Urea, Lactodifucotetraose, Trimethylamine, L-carnitine & Choline

Using this Method it is possible to analyze Compounds present in a Human Plasma Samples. The mechanism of retention of these polar compounds on the Column used is unknown at this time.

It is possible to Retain and Separate steroids and other Compounds based on the Shape of the Molecules.

Peaks:
1. Urea 61.0396 m/z [M + H]+
2. Lactodifucotetraose 635.2393 m/z [M + H]+
3. Trimethylamine N-oxide 76.0757 m/z [M + H]+
4. L-carnitine 162.1125 m/z [M + H]+
5. Choline 105.1148 m/z [M + H]+

Method Conditions
Column: Cogent UDC Cholesterol™, 4μm, 100Å
Catalog No.: 69069-05P-2
Dimensions: 2.1 x 50mm
Mobile Phase:
A: DI Water / 0.1% Formic Acid (v/v)
B: Acetonitrile / 0.1% Formic Acid (v/v)
Gradient:

Time (minutes) %B
0 90
1 90
7 20
11 20
12 90

Post Time: 3 minutes
Flow rate: 0.4 mL/minute
Detection: ESI – POS – Agilent 6210 MSD TOF Mass Spectrometer
Injection vol.: 1μL
Sample Preparation: Proprietary Lyophilized Plasma Sample was reconstituted in 60 µL 80% Acetonitrile / 20% DI Water mixture.

Attachment