MicroSolv Technologies Knowledge Base

Microsolv

Browse Ask a Question
 
Tools
Categories

How to Condition Cogent TYPE-C Silica™ HPLC Columns for Storage

Reference Number: AA-01569 Created: 06/22/2013 02:03 PM Last Updated: 05/22/2019 02:55 PM

This article describes how to prepare the Cogent TYPE-C Silica™ columns for initial use when you first receive them as well as how to store them for either overnight or long-term to prevent column degradation.

Solvents to be used

Storage Solvents for Reverse Phase (RP) or Aqueous Normal Phase (ANP):
If RP and ANP are your desired next use for these columns, storage in a mixture of 20% Distilled Water that is acidified with 0.1% v/v formic acid and 80% Acetonitrile is recommended. If required, 100% Acetonitrile may be used.

Storage Solvents for Organic Normal Phase:
If ONP is your desired next use for these columns, storage in hexane is recommended.

Note:
Due to the immiscibility of many organic normal phase (ONP) solvents and aqueous NP/RP based solvents, conversion from ONP to ANP/ARP can be best effected by flowing a mutually miscible solvent, such as methylene chloride, through the column for 15 bed volumes, prior to re-equilibration in the aqueous or organic based solvents to be next used. A simple 30 minute procedure (see below) allows switching from one mode to another.


Procedure:
A – for moving from RP-HPLC to NP-HPLC pump 100% methanol for 15 minutes at 1 mL/min. flow rate, followed by 15 minutes 100% methylene chloride. Column is ready to be equilibrated with mobile phase for NP-HPLC.
B – for moving from NP-HPLC to RP-HPLC pump 100% methylene chloride for 15 minutes at 1 mL/min. flow rate, followed by 15 minutes 100% methanol. Column is ready to be equilibrated with mobile phase for RP-HPLC.

       Due to the minimal water bonding to the TYPE-C Silica™ particle surface, bonded and un-bonded TYPE-C™ columns equilibrate from organic/aqueous and to pH changes much quicker than most irregular, type-A or type-B based columns.

Important:
•  Store column at room temperature. Never refrigerate or store in heated appliances.
•  Never let the column dry out.
•  Never store the column intended for use in RP HPLC in solvents containing DI water without the suggested 0.1% formic acid below.



Storage Procedures


How to Store: Cogent TYPE-C™ HPLC columns for Reversed Phase or Aqueous Normal Phase use (ANP).

Overnight:
1.Rinse the column with 8-10 column volumes with a mixture of 20% distilled water that is acidified with 0.1% v/v formic acid and 80% acetonitrile. If required 100% acetonitrile can be used instead.
   Cap the ends of the column with the protective end caps or column plugs provided. Leaving ends of the column open will cause permanent damage to the column.

Long Term Storage:
1.Rinse the column with 15 – 20 column volumes of 20% distilled water that is acidified with 0.1% v/v formic acid and 80% acetonitrile. If required 100% acetonitrile can be used instead.
   Cap the ends of the column with the protective end caps or column plugs provided.
2.Return column to the box it came in for safekeeping.
3.Store column at room temperature. Do not store in refrigerators or in heated appliances.

** Diamond Hydride™ columns have a special storage protocol, given here.



How to Store Cogent TYPE-C™ HPLC columns for Organic Normal Phase (ONP) use.

Overnight:
1.Rinse the column with 8-10 column volumes of 100% Hexane or other non polar solvent.
2.Cap the ends of the column with the protective end caps or column plugs provided. Leaving ends of the column open will cause permanent damage to the column.

Long Term Storage:
1.Rinse the column with 15 – 20 column volumes of 100% hexane or other non polar solvent.
2.Cap the ends of the column with the protective end caps or column plugs provided.
3.Return column to the box it came in for safekeeping.
4.Store column at room temperature. Do not store in refrigerators or heated appliances.
5.MARK THE COLUMN AS DESIGNATED FOR ORGANIC NORMAL PHASE AND WRITE WHAT SOLVENT IS INSIDE.

   
     © Copyright MicroSolv Technology Corporation, all rights reserved.


Info Ask a Question 