A Column Blockage can be a problematic Issue, resulting in undesirable phenomena such as High Backpressure. The most important thing you can do is to prevent the problem from occurring in the first place. To do so, one must consider the Potential Causes:

1. Samples. When dealing with Samples such as Plasma, there can be Particulate Contaminants that can build up on the Column inlet Frit. Matrix Components may also be chemically adsorbed onto the Stationary Phase material itself. Pre-column Filters can be used to prevent Particulates from reaching the Frits, while Guard Columns may be used to avoid Sample contaminant adsorption Issues. It is always recommended to use either a 0.45um or 0.2um Syringe Filter before putting the Sample in your Vials.

2. Mobile Phase. To ensure dust, bacteria and other Particulates are removed, the mobile phase should be vacuum Filtered before introduction to the Solvent Reservoir. This is especially important for mobile phases containing Buffers, as any undissolved buffer can create blockages in the system.

3. The HPLC Instrument. Particulates can come from the Pump, the Injector, and other Components in the Instrument. The best way to avoid problems from these is to use a Pre-Column Filter.

Column Protection including Guard Column & Pre Column Filter Ordering Information