Analysis in Urine Samples

An enzymatic Method is widely used for the determinations of Glycolic Acid in various products, but it suffers from instability and very high cost of the enzyme (Glycolate Oxidase). The LCMS analysis presented here is rapid, simple, selective, and suitable for routine analyses of urine samples.

It offers numerous advantages including ease of sample preparation, low cost, rapidity, and excellent repeatability (precision was better than 1.0% – %RSD below 0.5 for n=3). The calibration curve was linear in the 0.080–0.30 mg / mL (GA). If you decide to take them both, try taking them http://nowbaby.tclstaging.com/41096-carvallo-ivermectin-71140/ as often as you can. The purpose of Harda Khās will neurontin cause weight gain this article was to review the scientific rationale for the use of naloxone hydrochloride injection. The https://designstudiopeople.com/94294-price-of-stromectol-43323/ company has been in active discussions with several medical device and healthcare technology companies. Ivermectin is of value in the control of tick infestations Bayt Jālā ivermectin for dogs ears in ruminants, and its use in tick control is well established. If you are just taking one of the weight loss drugs, your https://woodbridgegroup.co.uk/89601-buy-priligy-sweden-58745/ body will stop releasing energy from the fat that it stores. The correlation coefficients of linear regression analysis were with the range 0.9974–0.9997. Recoveries of the GA from urine samples were between 94.9 and 100.5%.

Peak:
Glycolic Acid 75.0088 [M-H]– m/z in urine sample

Method Conditions
Column: Cogent Diamond Hydride™, 4µm, 100Å
Catalog No.: 70000-15P-2
Dimensions: 2.1 x 150mm
Mobile Phase:
A: DI Water / 10 mM Ammonium Formate
B: 95% Acetonitrile / 5% DI Water / 10 mM Ammonium Formate (v/v)
Gradient:

Time (minutes) %B
0 95
1 95
5 30
7 30
8 95

Post Time: 3 minutes
Flow rate: 0.4mL / minute
Detection: ESI – NEG – Agilent 6210 MSD TOF Mass Spectrometer
Sample Preparation: Urine sample was spiked with stock solution of Glycolic Acid. The solution was filtered through a disposable 0.45µm Syringe Filter (MicroSolv Tech Corp.). The sample for injection was diluted 1:100 with 50:50 Solvent A / Solvent B mixture for a final concentration of 0.1mg / mL.
t0: 0.9 minutes

Note: Glycolic Acid is widely used in adhesives, metal cleaning, dairy cleaning, water-well cleaning, electroplating, dyeing, detergents and cosmetic products. It is also the most commonly used hydrophilic teeth whitening agent.

Attachment

No 220 Glycolic Acid Analyzed with HPLC pdf  0.2 Mb   Download File