Rugged & Fast Without Paired Ion Reagents
Ion-pair Chromatography (IPC) is commonly used in order to retain Atropine on ordinary Reversed Phase HPLC Columns. Beside long equilibration times with these columns, IPC often suffers from poor robustness.
Using an isocratic elution with flow rate 1mL / minutes gave symmetrical peak for this Tropane Alkaloid. The sensitivity is sufficient for the proper determination of Atropine in plasma after intravenous administration of the drug to hospitalized patients.
This method is also useful for testing for drug poisoning and for stability testing of Atropine solutions during manufacturing. The linear range of detection for Atropine was around 5.0µg / mL with a limit of quantification (LOQ) 10.0µg / mL.
Peak:
Atropine (RT = 5.94 min)
Method Conditions
Column: Cogent UDC-Cholesterol™, 4µm, 100Å
Catalog No.: 69069-15P
Dimensions: 4.6 x 150mm
Mobile Phase:
—A: DI Water + 0.1% Acetic Acid + 0.005% TFA
—B: Acetonitrile + 0.1% Acetic Acid + 0.005% TFA
Flow rate: 1mL / minute
Detection: UV @ 214nm
Injection vol.: 1μL
Sample Preparation: Prepared in 50% Solution A / 50% Solution B, concentration 1mg / mL Sample was filtered through a 0.45µm Nylon Syringe Filter prior to HPLC-UV injections (MicroSolv Technology Corp.)
Note: Being potentially deadly, Atropine derives its name from Atropos, one of the three Fates who, according to Greek mythology, chose how a person was to die. Atropine is a core medicine in the World Health Organization’s “Essential Drugs List”, which is a list of minimum medical needs for a basic health care system.
Attachment
No 119 Determination of Atropine with HPLC pdf 0.1 Mb Download File