Excellent Peak Shape for an Organic Amine
This Assay is easy to perform with a well-retained analyte Peak. Since Peak Tailing can often be an issue with organic amines, such as this compound, using conventional HPLC Columns it is notable that this Method produces a very Symmetrical Peak.
In the Chromatogram below, an overlay of injections from two lots of Columns is shown to illustrate Robustness and Precision of the Method which is suitable for routine assay of Metoprolol Tartrate Formulations.
Peak:
Metoprolol
Method Conditions
Column: Cogent Diamond Hydride™, 4μm, 100Å
Catalog No.: 70000-7.5P
Dimensions: 4.6 x 75mm
Mobile Phase:
—A: DI Water with 0.1% Trifluoroacetic Acid (TFA) v/v
—B: Acetonitrile with 0.1% Trifluoroacetic Acid (TFA) v/v
Gradient:
Time (minutes) | %B |
0 | 95 |
1 | 95 |
6 | 50 |
7 | 95 |
Post Time: 3 minutes
Injection vol.: 1μL
Flow rate: 1.0mL / minute
Detection: UV @ 215nm
Sample Preparation: 1mg Metoprolol Tartrate USP Reference Standard was dissolved in 1mL of 50:50 Solvent A / Solvent B. This stock solution was diluted 1:10 for HPLC injections using the same diluent.
t0: 0.9 minutes
Note: Metoprolol is a selective ß1 receptor blocker. It is used to treat various cardiovascular conditions such as hypertension. Various trade names for the drug are available such as Lopressor® and Toprol®. It is sometimes found in combination formulations with hydrochlorothiazide.
Attachment
No 208 Metoprolol Tartrate Analyzed with HPLC pdf 0.4 Mb Download File