API and Related Compounds Separation

Furazolidone was Separated from related compounds in a commercially available pharmaceutical preparation. Excellent Peak shape and Selectivity from the related compounds was achieved with this Robust Method making it a great choice for QC applications even when using aggressive Mobile Phases that include acids. Molecular weight of Furazolidone is 225.16.

Peak:
Furazolidone

Method Conditions
Column: Cogent Bidentate C18™, 4µm, 100Å
Catalog No.: 40018-15P
Dimensions: 4.6 x 150mm
Mobile Phase: 12% Acetonitrile / 88% DI Water / 1% Acetic Acid
Temperature: 30°C
Injection vol.: 20µL
Flow rate: 2mL / minute
Detection: UV @ 365nm

Note: This antibiotic drug is used to treat diarrhea and enteritis caused by bacteria or protozoan infection. Also used as a veterinary medicine for salmonids and Myxobolus cerebralis infection.

Attachment

No 73 Furazolidone Analyzed with HPLC pdf 0.2 Mb  Download File