Date: 29-JULY-2020 Last Updated: 17-JUNE-2025
Answer: Not all three ranges.
The MICROSOLV UV-Vis Calibration Kit™ is designed to support calibration in the UV (200–400 nm) and Visible (400–780 nm) spectral regions. It includes:
- Potassium dichromate for absorbance accuracy
- Holmium oxide for wavelength accuracy
- Potassium iodide for stray light assessment
These components are suitable for compliance with USP <857>, but only within the wavelength ranges covered by the included solutions.
This kit does not support calibration in the Near-Infrared (NIR) range, which typically spans 780–3300 nm, depending on the spectrophotometer. NIR calibration requires different standards and procedures not included in this kit.
For reference, the chart below outlines the acceptable criteria and wavelength ranges defined in USP <857>. The highlighted section indicates the specific regions addressed by our calibration kit.\