Does the UV-vis calibration kit operate in all three ranges UV and visible and visible/NIR - FAQ
July 29, 2020
/
/
/

Date: 29-JULY-2020   Last Updated: 17-JUNE-2025

Answer: Not all three ranges.

The MICROSOLV UV-Vis Calibration Kit™ is designed to support calibration in the UV (200–400 nm) and Visible (400–780 nm) spectral regions. It includes:

  • Potassium dichromate for absorbance accuracy
  • Holmium oxide for wavelength accuracy
  • Potassium iodide for stray light assessment

These components are suitable for compliance with USP <857>, but only within the wavelength ranges covered by the included solutions.

This kit does not support calibration in the Near-Infrared (NIR) range, which typically spans 780–3300 nm, depending on the spectrophotometer. NIR calibration requires different standards and procedures not included in this kit.

For reference, the chart below outlines the acceptable criteria and wavelength ranges defined in USP <857>. The highlighted section indicates the specific regions addressed by our calibration kit.\

 