Since Cogent TYPE-C columns can be used in Reversed Phase or Normal Phase what do you suggest to switch these solvent systems which are not miscible - FAQ
April 14, 2020
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/

Suggested Procedure:

A – Changing from Reversed Phase to Normal Phase HPLC; pump 100% methanol for 15 minutes at 1 mL / minute flow rate, followed by 15 minutes 100% methylene chloride. The column is ready to be equilibrated with mobile phase for NP HPLC.

B – Changing from Normal Phase to Reverses Phase HPLC; pump 100% methylene chloride for 15 minutes at 1 mL / minute flow rate, followed by 15 minutes 100% methanol. The column is ready to be equilibrated with mobile phase for RP-HPLC.

A simple 30 minute procedure allows switching from one mode to another.

© Copyright 2025. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media