Can the HSQ Kit be used to measure the LOQ and LOD differences between instruments - FAQ
April 7, 2012
/
/
/
/
/
/
/
/
/
/
/
Yes, this is another really great “extra feature” of the HSQ Kit™.
By determining the, signal to noise ratio, the extra column volumes and the dwell volume of each instrument using the HSQ Kit™ and the software provided, you can determine the differences in LOQ (limits of quantitation) and LOD (limits of detection) between instruments. This can be very helpful in troubleshooting lack of precision in methods that transfer from one instrument to the other.
Click HERE for HSQ Kit ordering information and pictures.
By determining the, signal to noise ratio, the extra column volumes and the dwell volume of each instrument using the HSQ Kit™ and the software provided, you can determine the differences in LOQ (limits of quantitation) and LOD (limits of detection) between instruments. This can be very helpful in troubleshooting lack of precision in methods that transfer from one instrument to the other.
Click HERE for HSQ Kit ordering information and pictures.