Reference values for holmium oxide other than 1 nm – Tech Information
August 7, 2024
/
/
/
/
/
/
/
/
When dealing with reference values (nm) that aren’t explicitly listed as bandwidths, you have a couple of options:
a. For information only. Cell = 1.0 cm; Ref = H2O
Click HERE for PQ Kits for Ordering Information. See the NIST reference Standard Reference Material® 2034 HERE.
- Interpolation: You can interpolate between the provided values. This means estimating the value based on the known data points. It’s a useful approach when you need a precise value within the given range.
- Next Value Approximation: Alternatively, you can use the next higher value as an approximation. Since the official values typically vary only in the third or fourth significant figure, this slight difference won’t be detectable given the instrument’s accuracy of 1-2 nm.

a. For information only. Cell = 1.0 cm; Ref = H2O
Click HERE for PQ Kits for Ordering Information. See the NIST reference Standard Reference Material® 2034 HERE.