Fluctuations in readings when using the UV-Vis Calibration Kit - Tech Information
April 3, 2013
/
/
/
/
/
/
/
/
/
Fluctuations in readings in a UV-vis instrument are almost certainly due to a failing lamp. 

A typical instrument which uses a xenon lamp is normally rated for only about 500 hours, while a deuterium can range from 1,000-3,000 hrs.  However, they will sooner or later need to be replaced.  If you are not sure of the hours on the lamp, I would start by replacing it to eliminate any fluctuations. If changing the lamp doesn't fix the problem, then you may need a service call.

Using the UV-Vis Calibration Kit will flag a problem such as this before erratic or poor data results from the bad lamp.

Click HERE for UV-Vis calibration kit ordering information and pictures.