Cogent Diamond Hydride 2.o HPLC column retention time precision - AppNote
November 13, 2015
/
/
/
/
/
This graph is from a study investigating the day to day retention time precision with two test solutes on the Cogent Diamond Hydride 2.o™ HPLC column


Peaks:
Blue Trace: Phenylalanine
Red Trace: Phenylglycine

Many factors can affect column robustness and lifetime, so it is not possible to define specific values for those parameters.

Mobile Phase:
--A: DI Water / 0.1% Formic Acid
--B: Acetonitrile / 0.1% Formic Acid
Flow Rate: 0.3mL / minute
Injection Volume: 0.2uL
Detection: UV 254nm


Gradient:
Time (minutes) %B
0 90
6 20
7 20
8 90
 



Cogent Diamond Hydride HPLC Column Ordering Information

Related Articles

  1. What pressure should I expect with a Diamond Hydride 2.o HPLC column at flow rates from 0.2 to 2.0mL / min - FAQ

© Copyright 2025. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media