Standard silicone pre slit sealing mat compared to solvent resistant mat - FAQ
November 7, 2020
/
/
/
/
/
/
/
Comparison of 96 well sealing mats, silicone rubber v. chemical resistant U-2D mats.

Silicone rubber mats should only be used with pure aqueous and buffer solutions.

The solvent resistant mat can be used with methanol, acetonitrile and water / buffers. These should not be used with non-polar solvents or aggressive acid or base such as ethyl acetate.

U-2D Silicone Mats Product Page

Related Articles

  1. Can an autosampler needle pierce the Zone Free sealing film and will it reseal afterwards - FAQ
  2. Which type of vial do you recommend to use with the U-2D Micro Sample Management inserts - FAQ