HPLC Column, RP Phenyl, Wide Pore, 10um, 4.6mm ID x 100mm Length, 300A. Cogent Brand. 1 EA.

Additional Information:

Cogent RP Phenyl Wide Pore is MICROSOLV's most popular phenyl stationary phase for reversed-phase HPLC and LC-MS applications involving larger molecules. Manufactured with high-purity silica and a C3-phenyl bonded phase featuring exhaustive double endcapping, this column provides balanced hydrophobic retention combined with strong π-π interactions. The result is enhanced selectivity for aromatic, heteroaromatic, and moderately polar compounds that may not separate adequately on traditional C18 or C8 columns. The 300 Å wide pore structure improves accessibility for larger molecules while maintaining the aromatic selectivity that makes RP Phenyl phases a preferred choice for many challenging separations. This stationary phase is also suitable as an alternative for USP L11 methods.

68024-10P-3M is an analytical HPLC column with dimensions of 4.6 mm ID x 100 mm length, packed with 10 µm RP Phenyl Wide Pore particles featuring a 300 Å pore size. The 100 mm column length provides increased resolving power compared to shorter columns while maintaining efficient analysis times. The C3-phenyl bonded phase provides improved selectivity for aromatic compounds through π-π interaction mechanisms while maintaining reproducible retention, excellent peak symmetry, reduced peak tailing, and lower silanol activity. The 10 µm particle size supports higher flow rates while generating lower backpressure than smaller particle columns. Typical flow rates are approximately 1.0–2.0 mL/min depending on mobile phase composition and system configuration.

This column is commonly used for aromatic pharmaceuticals, positional isomers, peptides, tocopherols, flavonoids, polynuclear aromatic compounds, nitro-aromatic compounds, APIs, natural products, and related compounds. The wide pore architecture can offer advantages for larger molecules while preserving the unique phenyl selectivity. It is particularly useful when structurally similar analytes co-elute on C18 columns or when additional aromatic selectivity is desired. The 100 mm length provides enhanced resolution for method development, QC testing, and routine analytical applications while maintaining compatibility with preparative method scouting workflows.

Key Benefits

  • RP Phenyl Wide Pore stationary phase with C3-phenyl bonded chemistry

  • 300 Å pore size for enhanced accessibility of larger molecules

  • Strong π-π interaction selectivity

  • Fully double endcapped

  • High-purity silica support

  • Reduced silanol activity

  • Reduced peak tailing

  • Excellent peak symmetry

  • Suitable alternative for USP L11 methods

  • Compatible with HPLC and LC-MS

Typical Applications

  • USP L11 methods

  • Positional isomer separations

  • Aromatic pharmaceutical compounds

  • Active pharmaceutical ingredient (API) analysis

  • Peptide separations

  • Tocopherol analysis

  • Flavonoid analysis

  • Polynuclear aromatic compound analysis

  • Nitro-aromatic compound analysis

  • Natural product analysis

Specifications

  • Product Type: HPLC Column

  • Stationary Phase: RP Phenyl Wide Pore

  • Chromatography Mode: Reversed Phase

  • USP Classification: L11

  • Bonded Phase: C3-Phenyl (Propyl Phenyl)

  • Endcapping: Fully Double Endcapped

  • Particle Size: 10 µm

  • Pore Size: 300 Å

  • Column Dimensions: 4.6 mm ID x 100 mm

  • Length: 100 mm

  • Typical Flow Rate: 1.0–2.0 mL/min

  • Brand: Cogent

Additional Resources

Important Notes

  • Suitable as an alternative for USP L11 methods

  • Often selected when C18 or C8 columns do not provide sufficient selectivity

  • Excellent for aromatic and structurally similar compounds

  • Effective for peptides and moderately polar analytes

  • Wide pore design can improve accessibility for larger molecules

  • Commonly used in GMP, QC, and research laboratories

  • Enhanced resolution compared to shorter column formats

  • Frequently used for preparative method scouting and method development

  • A Cogent Column Filter is included and recommended to help protect the column and extend service life


 

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media