Organic acids and anions play an essential role in both environmental and biological processes. Determination of these molecules is important for both quality control and research laboratories. This application note shows the efficiency of CElixirOA™ to separate anions and organic acids at pH 5.4.

Method Conditions:
Voltage: 30kV
Capillary: MicroSolvCE™ Bare Fused Silica, 75µm x 60cm
Injection: 5 sec pressure
Run Buffer: CElixirOA™ pH 5.4 Kit
Detection: Indirect UV 233nm
Polarity: Reverse
Standards Concentration: 20ppm

Method:
Using a Beckman P/ACE MDQ™ HPCE System and 75µm ID bare fused silica capillary, the analysis and separation of over 10 anions and organic acids is easily achieved. The 1.25g of Sugar Molasses was diluted in 100ml of NaOH 0.05M and was hydro-dynamically injected for 5 seconds. The capillary was conditioned according the CElixirOA™ manual. The separation was completed in less than 7 minutes. Please note that in reverse polarity, the sample is introduced at the cathode and migrates to the anode.

Discussion and Rationale:
Many organic acids and small anions are UV transparent, therefore a direct detection method is not usable. CElixirOA™ pH 5.4 Kit contains pyridine-dicarboxylic acid (background electrolyte) and a chromophore that absorbs UV energy at 233nm. Using the dynamic coating technology of CElixir with a polycation and the included chromophore, the separation shown is possible and easy with no hassles of mixing solutions or preparing any run buffers on your own.

Comments:
CElixirOA™ provides simple, reliable and rapid separations covering a wide range of applications. With good linearity and reproducibility below 0.5% RSD for migration times (click here to view chart) CElixirOA™ is a valuable alternative to traditional techniques such as Ion Chromatography.