Separation of Valaciclovir from two Degradants

The anti-viral Herpes drug Valaciclovir and its degradation products are analyzed using Aqueous Normal Phase HPLC with MS detection. The Method development was completed using UV detection, where excellent Separation between the Prodrug Valaciclovir and its active converted degradant, Acyclovir, was observed.

The Method transfer from UV to MS detection was accomplished using only a change of the flow rate. When MS detection was used, a second Degradant (D,L-Valine, formed from hydrolysis of the prodrug) was also detected without derivatization.

Peaks:
Figure A: Valaciclovir 325.1619 m/z
Figure B: Acyclovir 226.0935 [M+H]+ (degradant / active form)
Peaks 1 and 1′ have the same mass. 1′ is the Peak that corresponds to the Acyclovir formed in the source
Figure C: D,L-Valine 118.0863 m/z [M+H]+

Method Conditions
Column: Cogent Diamond Hydride™, 4µm, 100Å
Catalog No.: 70000-15P-2
Dimensions: 2.1 x 150mm
Mobile Phase:
A: DI Water with 0.1% Formic Acid (v/v)
B: Acetonitrile with 0.1% Formic Acid (v/v)
Gradient:

Time (minutes) %B
0 95
1 95
6 40
7 40
8 95

Post Time: 3 minutes
Injection vol.: 1µL
Flow rate: 0.4mL / minute
Detection: ESI – POS – Agilent 6210 MSD TOF Mass Spectrometer
Sample Preparation: Stock Solution: 1000mg strength Valtrex® tablet was ground and added to 100mL volumetric flask containing 50mL 50:50 DI Water / Acetonitrile diluent. The solution was sonicated 10 minutes, diluted to mark, and mixed. A portion was filtered through a 0.45µm Nylon Syringe Filter (MicroSolv Tech Corp.). The stock solution was diluted 1:100 with 50:50 Acetonitrile / DI Water (Figure A). Another aliquot was diluted 1:100 with 50:50 1N HCL / Acetonitrile mixture and heated at 85°C for 30 minutes (Figures B and C).
t0: 0.9 minutes

Attachment

No 256 Valaciclovir & Degradants Analyzed with LCMS pdf 0.3 Mb  Download File