Tip on how to calculate extra-column volume using the MicroSolv HSQ™ Kit software

Reference Number: AA-02458 Created: 10/14/2014 11:17 AM Last Updated: 11/25/2014 11:36 AM

The extra-column volume test is calculated from the single injection of the Resolution Test Mixture.  It uses the RTM method, which is simply the basic PQ™ method, but with a runtime of 3 minutes.  The other difference in the RTM method is that is used to acquire the caffeine spectrum, if it is a DAD detector, so in general the bandwidth would be set narrower for improved spectral resolution, even though more noise would be produced. 
 
The calculations of the extra-column dispersion are detailed if one prints out the equations.  In an ideal system, all peaks in the chromatogram would have nearly the same efficiency.  Extra-column dispersion influences the early peaks more than the later ones, since it is a larger percentage of the overall variance.  Thus, the plate counts tend to steadily increase to a constant level with longer retention.  The extra-column variance is calculated from those data.  Also note that it is not really extra-column volume that is measured, but rather extra-column volume variance (or after the square root, extra-column volume standard deviation).  Most analytical HPLC’s will produce about 10uL of ec volume standard deviation dispersion (or 100 uL^2 of volume variance), while UPLC’s should produce less than half that value if they are properly plumbed.

    
     HSQ™ Kit Product Page
     Download latest HSQ™ software