How to change ferrules on HPLC column connectors

Reference Number: AA-00416 Created: 04/13/2012 08:54 PM Last Updated: 04/13/2012 08:56 PM

Frit Replacement HPLC Column Connectors
   So Easy...
       Anyone Can Use It!



Ferrule Replacement Instructions for the HPLC Column Connectors

  1. Change ferrule after approximately every 50 insertions or when excessive ferrule wear is detected.
  2. To change ferrule, grasp ferrule in the area of the groove with a pair of wire cutters using minimal pressure and gently pull ferrule from the threaded stem, wiggling the ferrule if necessary.
  3. Insert a new ferrule into the threaded stem and screw connector into a port to seat the ferrule into the threaded stem.

The HPLC Column Connectors are designed to provide universal connection to any manufacturer’s 1/16th 10-32 port. Maximum pressure, 6,000 psi.